Handling and Storing

Observe the warnings and cautions in this section when handling and storing BK Medical transducers and attachments.

Download Care, Cleaning and Safety for more detailed information.

Handling Transducers

Transducers are designed for durability but must be handled with care. Dropping or bumping them can cause serious damage. Don’t step on the cord, or run over it with the wheels of the scanner. This damage is not covered by the warranty or the service contract.

WARNING To maintain safe operation, the transducer plug must always be completely dry.

Storing Transducers When Not in Use

Transducers should always be stored in a safe place. Do not leave them lying on a table, for example. They should not be stored with a watertight plug lid or cover screwed on.

When you are not using a transducer, place it in one of the transducer holders on the scanner.

Operating and Storage Environment

Environmental limits for BK Medical scanners during operation and storage. Environmental limits during disinfection processing are also shown.

Environmental Limits Maximum Minimum
Storage temperature +70°C (+158°F) -25°C (-13°F)
Storage humidity 85% RH
Operating temperature +40°C (+104°F) +10°C (+50°F)
Operating pressure 1060hPa (15.4psi) 700hPa (10.2psi)
Watertight immersion temperature +40°C (+104°F) NA (not applicable
Watertight immersion time 15 hours per 24 hours NA
Pressure during gas processing NA 500hPa (7.3psi)
Pressure during gas processing for 8666, 8814, 8815, and 8816 transducers NA 100hPa (1.5psi)
Temperature during gas processing 55°C (+131°F) NA